Table 5 Mean amplitude at peak frequency measured with each consumer product accelerometer (CPA) and laboratory grade accelerometer (LGA) as reference in extended arm position based on 20 measurements

From: Quantification of tremor using consumer product accelerometry is feasible in patients with essential tremor and Parkinson’s disease: a comparative study

Location

Device

Mean (mG^2/Hz)

Intra-individual variability

Coefficient of variation (%)

Inter-individual

Coefficient of variation (%)

Distal

Reference LGAa

38.27

27.91

322.91

Distal

Reference LGA b

20.31

37.26

316.00

Distal

iPhone 7

149.77

44.39

328.17

Distal

iPod Touch 5

869.93

51.00

403.96

Distal

Huawei Nexus 6P

666.42

66.79

407.66

Distal

MW Clip

1656.51

90.70

400.59

Proximal

Apple Watch 2

31.62

42.57

376.02

Proximal

Huawei Watch

67.34

124.33

431.74

Proximal

MW Watch

34.24

37.74

374.98

  1. aReference LGA measurement prior to CPA hand-held devices bReference LGA measurement prior to CPA wrist-worn devices