Table 4 Mean amplitude at peak frequency measured with each consumer product accelerometer (CPA) and laboratory grade accelerometer (LGA) as reference in resting arm position based on 20 patients

From: Quantification of tremor using consumer product accelerometry is feasible in patients with essential tremor and Parkinson’s disease: a comparative study

Location

Device

Mean (mG^2/Hz)

Intra-individual variability

Coefficient of variation (%)

Inter-individual variability

Coefficient of variation (%)

Distal

Reference LGAa

794.58

88.17

293.11

Distal

Reference LGAb

384.21

89.47

342.36

Distal

iPhone 7

204.92

26.78

345.72

Distal

iPod Touch 5

706.81

45.08

311.90

Distal

Huawei Nexus 6P

832.41

54.85

307.80

Distal

MW Clip

5061.86

41.69

306.48

Proximal

Apple Watch 2

257.64

57.28

370.40

Proximal

Huawei Watch

341.01

52.58

383.31

Proximal

MW Watch

253.96

26.62

325.83

  1. aReference LGA measurement prior to CPA hand-held devices bReference LGA measurement prior to CPA wrist-worn devices