Table 3 Mean peak frequency L1 norm measured with each consumer product accelerometer (CPA) and laboratory grade accelerometer (LGA) as reference in extended arm position based on 20 patients

From: Quantification of tremor using consumer product accelerometry is feasible in patients with essential tremor and Parkinson’s disease: a comparative study

Location

Device

Mean Peak Frequency (Hz)

Intra-individual variability

Coefficient of variation (%)

Inter-individual variability

Coefficient of variation (%)

Distal

Reference LGAa

5.61

23.25

33.19

Distal

Reference LGAb

5.35

20.16

36.40

Distal

iPhone 7

5.48

10.95

28.07

Distal

iPod Touch 5

6.01

10.46

26.16

Distal

Huawei Nexus 6P

5.99

15.64

22.13

Distal

MW Clip

5.58

17.93

21.57

Proximal

Apple Watch 2

4.62

17.68

49.74

Proximal

Huawei Watch

4.68

25.42

44.09

Proximal

MW Watch

4.23

19.10

35.96

  1. aReference LGA measurement prior to CPA hand-held devices bReference LGA measurement prior to CPA wrist-worn devices