Table 2 Mean peak frequency L1 norm measured with each consumer product accelerometer (CPA) and laboratory grade accelerometer (LGA) as reference in resting arm position based on 20 patients

From: Quantification of tremor using consumer product accelerometry is feasible in patients with essential tremor and Parkinson’s disease: a comparative study

Location

Device

Mean Peak Frequency (Hz)

Intra-individual variability

Coefficient of variation (%)

Inter-individual variability

Coefficient of variation (%)

Distal

Reference LGAa

4.76

23.28

20.72

Distal

Reference LGAb

5.45

9.23

29.52

Distal

iPhone 7

5.29

8.66

23.19

Distal

iPod Touch 5

5.44

10.77

24.32

Distal

Huawei Nexus 6P

5.6

8.79

24.89

Distal

MW Clip

5.4

10.98

24.18

Proximal

Apple Watch 2

5.48

14.93

25.36

Proximal

Huawei Watch

4.81

18.45

32.27

Proximal

MW Watch

4.79

15.30

21.61

  1. aReference LGA measurement prior to CPA hand-held devices bReference LGA measurement prior to CPA wrist-worn devices